Minutes of the 2006/03/29 Blast analysis meeting

From: Michael Kohl (kohlm@mit.edu)
Date: Thu Mar 30 2006 - 23:52:23 EST


Hi,

we had a long meeting last Wednesday which was not always delightful.
Here's my recollection, I apologize for its length. Read it nevertheless.

Minutes:
-Find these minutes along with all plots shown in
http://blast.lns.mit.edu/PRIVATE_RESULTS/USEFUL/ANALYSIS_MEETINGS/meeting_060329/

-New spin angle measurements
  +Genya's plots (see excel file),
  +Chi's spin angle from ed elastic tensor asymmetry measurement was
   31.7 and 47.7 degrees for 2004 and 2005.
  +Chris' spin angle from ep elastic vector asymmetry measurement was
   47.5 degrees
  +ed-elastic yield-weighted average for January2005 map was 31.3 and
   46.8 degrees. ep-elastic yield weighted average was 47.1 degrees.
  +ed-elastic yield-weighted average for June2005 map according to CZ
   was 30.7 and 46.2 degrees, i.e. up to 1.5 degrees smaller than the
   angle determined from the asymmetry analysis.
  +Spin angle map with 3d hall probe has been repeated in March 2006.
   Comparison of June2005 map (Blast field on) with March2006 map
   (Blast field off): Profile is reproduced but shifted by ~1 degree in
   average, hence increasing the discrepancy with CZ to 2.5 degrees.
   Constancy of the difference along z could point to an alignment
   issue with the hall probe (would be good to check the constancy of
   the shift between two measured maps!!).
  +Effect of the Blast field on the March2005 map would decrease the
   angle by another ~0.4 degrees (still needs to be confirmed),
   increasing the discrepancy with Chi to as much as 3 degrees.
  +Results of the compass method are mostly (but not entirely) in
   agreement with the March2006 map.
  +Spin angle results with the long probe are ~0.5 degrees smaller than
   with the short probe.
  +Two sources of systematics for the compass method:
   1) The extension of the probe implies an averaging effect over the
      angle profile along the length. If the profile is an even
      distribution, a systematic shift would result.
   2) There could be a misalignment of the magnetic and geometrical
      axis of the probe.
  +Current strategy is to consider building a shorter probe whose
   magnetic axis can be adjusted to the geometrical axis at the same
   time.
  +To summarize, right now the problem is worse than before. We will
   discuss about the issue at the collaboration meeting next week.
  +My suggestion: The fact that the spin angle z-profiles are shifted by
   more or less constant amounts relative to each other raises doubts in
   the absolute alignment while relative measurements on the same jig are
   consistent. Hence, Chi's determination of the average spin angle from
   the tensor asymmetry analysis should be based on a comparison with
   the Montecarlo that uses a measured profile. (So far the used
   Montecarlo has assumed a flat distribution). Chi's subsequent
   analysis result for the average spin angle can then be compared with
   the ed-elastic yield weighted average of the initially used
   profile and will then reveal by how much the measured profile needs
   to be shifted in order to become realistic (=in agreement with the
   physics). The method needs to be cross-checked with the ep-elastic
   analysis (so far both ed and ep analyses have given consistent
   results). Result should be a corrected spin angle map that can be
   equally used by every analysis. Please let me know what you think.

-Mascarad+Epel issue
  +Mascarad only produces the radiative tail starting at a cutoff
   energy for the radiated photon (ad hoc set to 10 MeV).
  +Electron momentum generated with Mascarad is thus shifted relative
   to the unradiated momentum by at least 10 MeV.
  +Average momentum shift of electrons due to internal radiation
   convoluted with resolution can only be correctly estimated by
   Montecarlo if Mascarad is properly combined with the unradiated
   yield.
  +The proper combination of Mascarad with Epel needs to be established.

-Energy loss correction, kinematic corrections for v3_4_17
  +Eugene's plots (ppt).
  +Proton vs. electron angle shifts of 0.2 degrees within resolution of
   0.5 degrees
  +Introducing multiplicative offsets for momenta instead of shifts
  +Reconstructed beam energy from electron and proton variables. Energy
   loss effect for protons visible at large angles=low momenta. Beam
   energy from electron variables 10-20 MeV lower. Shift almost
   independent of the angle. Possible effect due to radiative-tail
   convoluted with resolution??

-Recipe for v3_4_17
  +For v3_4_17, kinematic corrections need to be produced and be made
   available (EG).
  +Electron radiative effect (convoluted with resolution) will be
   effectively absorbed in the residual kinematic correction for the
   electron.
  +Momentum loss for the proton should be parametrized and applied
   before residual kinematic corrections are determined.
  +Resulting kinematic corrections should be independent of energy
   loss, i.e. applicable to both proton and piplus data.
  +Usage of kinematic corrections by people also requires separately
   applying momentum loss correction.
  +Reconstructed Montecarlo for analysis should presently not be
   radiated, and should use energy loss correction for the proton.

-Geometry transformations
  +Chris' plots (gif)
  +Chris established and tested code that calculates derivatives of
   reconstructed variables with respect to shifts and rotations of the
   entire wire chamber in each sector.
  +Minimization procedure of p_e-p_e(th_e), p_p-p_p(th_p),
   th_e-the(th_p), phi_e-phi_e(phi_p) and z_e-z_e(z_p) for both sectors
   with a most likely position and orientation of the chambers.
  +Results so far in disagreement with range of variations allowed by
   the surveys.
  +Method needs some fine-tuning, some partameters may be more
   constrained than others, and the importance of the momentum, angle
   and vertex measurements has to be weighted with the respective
   resolutions.
  +Aki noted that the holding field was apparently not superponed with
   the Blast field for reconstruction (turned off in blastrc), which
   could be potentially responsible for the nonzero coplanarity.
   Will be addressed with the next recrunch.

-Timing calibration issue in 2005 runs in v3_4_17
  Chi's plots:
  +td-te shifted for runs 14336-14435; multiple peaks for all runs
  +copl: multiple peak + continuum, should be one peak
  +before: with static calibration no problem
  +diagnosis: T0's are wrong for left-14-top for 14228-~15200

  Investigation by MK (ppt)
  +Strategy of time calib was to use cosmics for time calib at a
   certain time and to use this also to calibrate the timing of the
   flasher signal such that the offsets found with the flasher
   run-by-run would be identical with the required offsets for physics
   run-by-run.
  +Flasher monitoring was problematic in 2005 because of periods with
   no flasher signals, laser replacement, and subsequent weak flasher
   amplitudes. Periods with missing flasher interpolated with constant.
  +Database monitor: Black is T0. Red is MeanTDC.
   Obviously wrong T0's for LT14T between 14336 and 15200. Why? T0's in
   database are differences between positions of flasher in TDC and
   flasher calib from cosmics. Flasher position in TDC = MeanTDC
  +Flasher calibration by cosmics done for 14134 before, and 16278 after
   -> huge difference for left-14-top, but also for l15t/b and r7t/b
   -little difference before and after laser replacement
  +Concluding, time calib with cosmics run 14134 wrong, to be fixed.
  +Calibration with cosmics has weakness that coplanarity is not well
   defined. Could be improved by comparing phi from tracking with phi
   from top-bottom timing.
  +Sidebands in flasher difference (with left-Tof-0-top subtracted).
   Mark had reported last September that in 6% of the runs timing is
   shifted. Problem due to peak finding failure!
    -> Will have to fix this run by run. Demonstrated that it can be
   fixed for the example of runs 16252-16254.
  +Any changes to the TOFs will possibly impact the calib for the
   neutrons, too.
  +Flasher resolution: Difference between two TOF channels is within
   2ns over months. Even static calib for T0's may work.
  +Motivation to use run-by-run calib was originally due to strongly
   varying ADC pedestals. Neutron timing relies on walk correction that
   requires proper ADC evaluation. It may be sufficient to assume
   constant T0's also for the neutron counters and only use run-by-run
   pedestals. However, simplest solution right now seems to me to fix
   the database for the wrong entries and stick with the same method.

-Conferences and contributions
List will be generated at collaboration meeting

Regards,

    Michael

+-------------------------------------+--------------------------+
| Office: | Home: |
|-------------------------------------|--------------------------|
| Dr. Michael Kohl | Michael Kohl |
| Laboratory for Nuclear Science | 5 Ibbetson Street |
| MIT-Bates Linear Accelerator Center | Somerville, MA 02143 |
| Middleton, MA 01949 | U.S.A. |
| U.S.A. | |
| - - - - - - - - - - - - | - - - - - - - - -|
| Email: kohlm@mit.edu | K.Michael.Kohl@gmx.de |
| Work: +1-617-253-9207 | Home: +1-617-629-3147 |
| Fax: +1-617-253-9599 | Mobile: +1-978-580-4190 |
| http://blast.lns.mit.edu | |
+-------------------------------------+--------------------------+



This archive was generated by hypermail 2.1.2 : Mon Feb 24 2014 - 14:07:33 EST